Model: ECG-603A |
|
Lead |
Standard 12 leads, simultaneous acquisition |
|
Input circuit |
Isolated input circuit with protection against pacemaker and defibrillator |
|
Input impedance |
>50MΩ |
|
Input circuit current |
<=0.1uA |
|
Calibration voltage |
1mV+/-3% |
|
Polarizing voltage |
+/-400mV |
|
Sensitivity |
5, 10, 20mm/mV, and automatic gain control |
|
Filter |
AC: 50Hz (-20dB); EMG: 35-45Hz (-3dB); DFT: <=0.45Hz |
|
Time constant |
3.2s (0, +20%) |
|
Noise level |
<=15uVP-P |
|
Frequency response |
0.05-150Hz(-3dB) |
|
CMRR |
>=100dB |
|
Sampling rate |
1ms |
|
A/D converter |
12 bit |
|
Recording paper |
Roll: 63mm*20m (30m) Zip fold: 63mm*75mm*20m |
|
Operation Mode |
Manual/Auto |
|
Printer |
High resolution thermal array printer |
|
Baseline Control |
Automatic elimination of baseline drift Automatic adjustment of baseline position |
|
Rhythm lead |
Selectable among standard 12 leads |
|
DC input |
10mm/v+/-5%; single ended input>=100ko 1V/mV+/-5%; single ended output>=100o |
|
Safety level |
Class I, Type CF |
|
Power requirements |
AC: 100-240V, 50/ 60Hz DC: Built in charger and Li-ion battery Supports continuous operation for more than |
|
Packing |
230mm×300mm×360mm |
|
Net/Gross Weight |
2.9kg/5.8kg |
Features
▲Hided portable and light weight. ▲Designed with a high-resolution thermal printer to print out ECG waveform, annotation as well related parameters for diagnosis reference. ▲Roll recording paper is 63mm in width and 30m in length. ▲Simultaneously acquisition of 12 lead ECG data, IEC Class, type CE safety standard. ▲Digital isolation technology is adopted to minimize time/temperature drift of components so as to promote adaptability of the unit to circumstance. ▲Digital filter for effecive inhibition of baseline drift, interference, as well automatic position of baseline ▲Self-test upon powering on. ▲Function of rhythm lead for observing abnormal ECG waveform & heart rate, Automatic measurement and interpretation of ECG waveform to relieve doctors from strain. ▲Multiple print formats include 3-channel, 2-channel, 1-channel, and particularly the one that combines 3-channel (for leadl-aVF) and 2-channel (for V1-V6). ▲AC/Rechargeable battery for supporting continuous examination when AC power supply is unavailable. ▲For battery operation, ECG-603A is equipped with a charger & system for battery capacity management and protection itself from excessive discharge. ▲Lfilters: AC, HUM, DFT Real-time clock, real-time measurement and presenting of ECG parameter. ▲Option: ECG software for net-work capability. ▲8 Patients Save / Copy / Communication ▲Large LCD to display 12 lead ECG waveform ▲Chinese / English language ▲Built in Rs232 interface (USB is optional)
|
|